
Water is present in tissues at concentrations 10,000 times higher than metabolites of interest (like NAA, Cho, Cr). An unprepared MR spectrum would thus be dominated by a giant water peak, while small organic molecules would be virtually undetectable above background noise.
To visualize these interesting small organic molecules, the large water peak must be suppressed. Methods to accomplish this are described below, the most common being CHESS (CHEmical Shift Selective saturation) or one of its variants.
CHESS was originally developed as a technique for fat suppression in conventional MR imaging, where it is commonly known by the generic name "fat-sat". By tuning the CHESS pulses to the resonant frequency of water instead of fat, water suppression may be obtained.
To visualize these interesting small organic molecules, the large water peak must be suppressed. Methods to accomplish this are described below, the most common being CHESS (CHEmical Shift Selective saturation) or one of its variants.
CHESS was originally developed as a technique for fat suppression in conventional MR imaging, where it is commonly known by the generic name "fat-sat". By tuning the CHESS pulses to the resonant frequency of water instead of fat, water suppression may be obtained.

A CHESS pulse selectively rotates water magnetization into the transverse plane where it is immediately dephased by application of a strong spoiler gradient. For MR spectroscopy a single CHESS pulse provides insufficient water suppression, so 3 CHESS pulses are used in the typical clinical implementation. To insure frequency selectivity, CHESS pulses are relatively long (20-30 ms). The optimal flip angle for each pulse depends on the interpulse spacing as well as local T1 and B1 effects. MR vendors offer automated water suppression procedures that iteratively evaluate and optimize flip angles based on the residual water signal.
Advanced Discussion (show/hide)»
At body temperatures the concentration of pure water in water is (¹H20 is 55.6 M. Because there are 2 hydrogens per water molecule, the concentration of ¹H in water is twice this much, or approximately 111 M. The fraction of water in brain tissue is about 70%, giving a water proton concentration of 0.70 x 111 ≈ 80 M.
Even with a superior level of water suppression in a perfectly shimmed magnet, some residual water resonance or secondary ripple effects are always present near δ = 4.65 ppm. Hence the spectral band from about 4.4 to 5.0 ppm remains largely uninterpretable for ¹H spectroscopy of tissues.
CHESS variants — all named with a "liquid" theme — include WET (Water suppression Enhanced through T1 effects), MOIST (Multiple Optimizations Insensitive Suppression Train), SWAMP (Suppression of Water with Adiabatic-Modulated Pulses) and VAPOR (VAriable Power radiofrequency pulses with Optimized Relaxation delays). These use 3, 4, 6, and 8 presaturation pulses respectively whose flip angles and interpulse delays have been optimized. While they offer improved water suppression over standard CHESS, especially for multi-voxel spectroscopy, they come with a time penalty. The 8-pulse VAPOR sequence, for example, requires up to 800 ms which increases the minimum sequence TR accordingly.
CHESS and its variants described above are all applied in the preparatory period prior to the MRS localization sequence. However, it is possible to incorporate frequency-selective suppression pulses in the middle of the spectroscopy acquisition. In STEAM, for example, a simple CHESS saturation pulse can be applied during the mixing interval (TM), since the at this time the magnetization is stored along the z-axis and is not undergoing T2-decay. For PRESS, either MEGA (MEscher-GArwood) or BASING (BAnd Selective Inversion with gradient dephasING) pulses with associated gradients can be interposed between the 180º-pulses. MEGA and BASING pulses can also be used for fat suppresison and spectral editing and will be described more completely in later Q&A's.
References
de Graaf RA, Nicolay K. Adiabatic water suppression using frequency selective excitation. Magn Reson Med 1998; 40:690–696. (First description of the SWAMP technique that uses adiabatic RF-pulses and is offered as a product by Philips).
Haase A, Frahm J, Hänicke W, Matthaei D. ¹H NMR chemical shift selective (CHESS) imaging. Phys Med Biol 1985; 30:341-344.
Mescher M, Tannus A, O’Neil Johnson M, Garwood M. Solvent suppression using selective echo dephasing. J Magn Reson A 1996; 123:226–229. (First description of the MEGA technique)
Mescher M, Merkle H, Kirsch J, et al. Simultaneous in vivo spectral editing and water suppression. NMR Biomed 1998; 11:266–272. (Further MEGA refinements)
Ogg RJ, Kingsley PB, Taylor JS. WET, a T1- and B1-insensitive water-suppression method for in vivo localized ¹H NMR spectroscopy. J Magn Reson Ser B 1994; 104:1-10.
Star-Lack J, Nelson SJ, Kurhanewicz J, et al. Improved water and lipid suppression for 3D PRESS CSI using RF band selective inversion with gradient dephasing (BASING). Magn Reson Med. 1997; 38:311–321.
Tkac I, Starcuk Z, Choi I-Y, Gruetter R. In vivo ¹H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med 1999; 41:649-656. (First description of the VAPOR technique).
de Graaf RA, Nicolay K. Adiabatic water suppression using frequency selective excitation. Magn Reson Med 1998; 40:690–696. (First description of the SWAMP technique that uses adiabatic RF-pulses and is offered as a product by Philips).
Haase A, Frahm J, Hänicke W, Matthaei D. ¹H NMR chemical shift selective (CHESS) imaging. Phys Med Biol 1985; 30:341-344.
Mescher M, Tannus A, O’Neil Johnson M, Garwood M. Solvent suppression using selective echo dephasing. J Magn Reson A 1996; 123:226–229. (First description of the MEGA technique)
Mescher M, Merkle H, Kirsch J, et al. Simultaneous in vivo spectral editing and water suppression. NMR Biomed 1998; 11:266–272. (Further MEGA refinements)
Ogg RJ, Kingsley PB, Taylor JS. WET, a T1- and B1-insensitive water-suppression method for in vivo localized ¹H NMR spectroscopy. J Magn Reson Ser B 1994; 104:1-10.
Star-Lack J, Nelson SJ, Kurhanewicz J, et al. Improved water and lipid suppression for 3D PRESS CSI using RF band selective inversion with gradient dephasing (BASING). Magn Reson Med. 1997; 38:311–321.
Tkac I, Starcuk Z, Choi I-Y, Gruetter R. In vivo ¹H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med 1999; 41:649-656. (First description of the VAPOR technique).
Related Questions
How do Fat-Sat pulses work?
How and why is fat suppression performed in an MRS study?
How do Fat-Sat pulses work?
How and why is fat suppression performed in an MRS study?